This is not a guide to how to collimate your telescope using a laser collimator or other method – see the links below for that.
To get the best out of your laser collimator you need to ensure it is correctly aligned.
When collimators are manufactured they have to be aligned, this means the laser beam has to be adjusted so it points exactly in the same direction as the body of the collimator. So when it’s inserted into your telescope the beam of light isn’t a lttle bit to the left, or right or up or down, it’s straight ahead and gives you accurate, consistent and useful results.
The cost of your collimator is a rough indication of how accurate it is, but even the best collimators can become misaligned and most of the worst collimators can be improved. This page will show you how.
Some assumptions
We’ll assume that the barrel of the collimator is uncomplicated so that if it is dropped into a channel and rotated it always points exactly at the same distant spot.
We’ll also assume that it has alignment screws that adjust the alignment of the internal laser. Often these screws are buried beneath rubber sealant which must be dug out.
Our last assumption is that you are able to set up the collimator so that it shines on a distannt wall to which you have fixed a sheet of white paper. The ideal distance is around 20m, but in a pinch even 10m will do.
Set up your channel. It must be solid, unmoving and it should hold the collimator firmly while allowing it to rotate. Some examples are shown below.
Shine the light on the distant surface. Turn the collimator to position 1. Positions 1, 2 and 3 correspond to the positions of the (usually) three grub screws that were used for adjustment at the factory before it was shipped out.
Rotate the collimator through positions 1 to 3, marking the position of the dot on your distance sheet of paper each time. Do this 2 or 3 times to confirm that the dot for each position is consistent. If it does not consistently return to its (1, 2 or 3) position each time then then that is a problem you must solve before moving on.
Once you have marked the position of the dot three times you have three points of a triangle. Mark the sides of the triangle and measure their lengths.
Measure the distance between the collimator and the sheet of paper.
To judge your collimator work out the smallest circle that would surround your three dots and use the table below to grade it from excellent to unusable.
| Distance collimator to sheet | Excellent (<2.5 arc min) | Good (< 5 arcmins) | Workable (<10 arcmins) | Unusable (>10arcmins) |
|---|---|---|---|---|
| 10m | <7mm | <15mm | <30mm | >30mm |
| 15m | <10mm | <20mm | <40mm | >40mm |
| 20m | <15mm | <30mm | <60mm | >60mm |
| 25m | <18mm | <35mm | <70mm | >70mm |
My low cost (less than £30) laser collimator started life as unusable and after one adjustment became workable. A second session produced a good collimator. Not sure though it will ever be excellent, but good is good enough.
Improving the collimator is a simple process of turning each alignment screw to move the dot closer to the centre of the white sheet, marking the three positions until you get as tight a grouping as your patience and tools allow.
DO NOT overtighten the screws are they kay strip their threads, be gentle. Remember that loosening2 screws can have the same effect as tightening that third screw.
Your aim from each session is to improve the collimator, not to make it perfect.
Good luck!